Moxi Z is the only automated cell counter that combines the Coulter Principle typically used in high-end cell counters with a patented thin-film sensor technology to allow for highly accurate (> 95%) and repeatable particle counting and sizing for a broad range of cell types – from mammalian cells to cells as small as wine yeast and more. Since today’s workflows demand accurate quality control of samples, determining cell counts precisely has a significant impact on outcomes and downstream costs.
This ultra-small instrument uses patented microfluidic thin-film cassettes that enable automatic load and measure operation. The resulting single cell volumetric measurements are completely technique-independent. It provides the ideal alternative to the tedious manual counting associated with cytometers, or the inaccurate results associated with image-based automated cell counters (typical accuracies of 75-80%).
Using the new Type S cassette, you can now accurately measure the smallest cells of any automated cell counter (down to 3µm average diameter) at concentrations of up to 2.5e+6/ml. This means the new Type S cassette can measure Yeast (including wine yeast) and non-spherical particles down to 14 fL in volume.
Download PDF:
- More accurate and repeatable than any other automated cell counter
- Results in just 8 seconds
- Ultra-small and ultra-easy to use (no manual counting, no focusing of any kind)
- Complete, high-resolution cell size/count histogram
- Handles “de-aggregation” and coincidence events
- Post-processing/analysis of count distributions with user-adjustable regions/gates
- Assess cell health (MPI) without the need for reagents like Trypan blue
- Complete data analysis and data portability (connectivity to PC, and MoxiChart software package)
- No maintenance required
Moxi Z | |
Resolution | 1200 histogram bins |
Weight | 1lb 7 oz |
Dimensions | 7.5″ L x 4.25″ W x 2.75″ H |
Battery | Lithium Ion 4500 mAh |
Data storage | 500 samples |
AC Power | 100-240V, 50/60 Hz, 0.2 Amps |
Connectivity | Mac/PC (Bluetooth and USB on-the-go) |
Data analysis | MoxiChart software package for Mac/PC |
Type M Cassette The Type M cassette is ideal for accurately measuring mammalian cells, or other particles 4-25 microns in average diameter. Since the technology is based on a volumetric measurement, non-spherical particles (34-8,180 fL) can also be measured accurately. | |
Dynamic range | 2 – 34 microns |
Average cell dia | 4 – 25 microns |
Average volume | 34 – 8,180 fL |
Sample concentration | 3,000 – 500,000 cells/mL |
Measurement time | 8 seconds |
Type S Cassette – NEW! The Type S cassette can accurately measure the smallest particles at concentrations of up to 2,500,000/ml. The Type S cassette is ideal for accurately measuring most yeast (including wine yeast), or other particles as small as 3 microns and up to 20 microns in average diameter. This includes some algae and protozoa. Since the technology is based on a volumetric measurement, non-spherical particles (14 – 4,200 fL) can also be measured accurately. | |
Dynamic range | 2 – 26 microns |
Average cell dia | 3 – 20 microns |
Average volume | 14 – 4,200 fL |
Sample concentration | 3,000 – 2,500,000 cells/mL |
Measurement time | 15 seconds |